Home » Projects » Vertisa Medical Waste Sterilization Devices PRE-500
Vertisa Medical Waste Sterilization Devices PROMED P500

Vertisa Medical Waste Sterilization Devices PRE-500

How does it function?

At the start of each cycle, when the upper lid opens, the operator loads the waste into the loading chamber.

Controlled by the PLC, the shredder adjusts its rotation direction to prevent blade jamming through its “anti-blocking” capability. Optionally, during the shredding process, the waste can be treated with microwaves generated by 6 integrated magnetron units.

Our uniquely designed shredder is proficient in grinding very hard materials such as surgical stainless steel instruments and hard ceramics.

Heating is accomplished by injecting saturated steam until the temperature is uniformly distributed in both the upper and lower chambers, reaching 134°C (274°F). Sterilization then occurs by maintaining a temperature of 134°C (274°F) to 138°C (280°F) and equivalent pressure for 10 to 15 minutes.

These conditions enable a reduction in pathogenic contamination levels in the medical waste by a factor of 8 Log 10.

During the cooling phase, any excess steam in the chamber is extracted, and the temperature of the ground sterilized waste is reduced by injecting cold air into the autoclave chamber.

The condensed water, which is sterile and safe for disposal, is drained from the system using either positive pressure or vacuum assistance, enabling it to be safely discharged into the sewage system.

Upon opening of the unloading lid and positioning a container beneath the waste drawer, the operator withdraws the drawer and empties the sterile waste into the container.

As the container is removed, the lower lid automatically closes, initiating the next cycle of the sterilizer.

PRE-500 Medical Waste Sterilization Device

  • Capacity: 175 kilograms per hour (385 pounds per hour)
  • Field of Utilization: Suitable for large-size hospitals and industrial sterilization facilities
  • Pressure Container: Constructed of high-quality stainless steel
  • Design and Production Certification: Certified by TUV Austria
  • Compliance: Produced in accordance with EU Directives PED (2014/68/EU), 2006/42/EC, 2014/30/EU & EU Norms EN285
  • Automation Control System: Sophisticated system with advanced reporting and monitoring software
  • Online Access: Supervision and troubleshooting available via internet connection
  • Fully Automatic Operation: No need for any operator intervention
  • Automatic Elevator System: Facilitates easy loading
  • Water Consumption: None
  • Fuel and Electrical Consumption: Low, thanks to integrated steam boilers equipped with state-of-the-art steam regeneration system.
Technical FeaturesValue
Size (L x W x H) (mm)2250 x 2250 x 4200
Weight (Kg)2500
Air Pressure (Bar max)8
Electrical connection required (Kw)19
Working CharacteristicsValue
Sterilizing Capacity (Kg/ Hour)150-175
Process Volume Capacity (Lt.)750
Average Waste Density (Kg/m3)100-150
Average Cycle Time (Min.)35-45
Maximum Steam Flow (Kg/Hour)320
Sterilization Efficiency (SAL)8Log10
Consumption / CycleValue
Steam (Kg)15
Electricity (Kw)5
Water (Lt)None
Special ConsumablesNone